Feyond Microplate Reader
Feyond series multi-mode microplate reader with absorbance, fluorescence, luminescence, TRF and FP detection for medical, biological and pharmaceutical research.
Send us an inquiry and our team will respond within 24 hours.
Feyond Series Multi-Mode Microplate Reader
The Feyond series multi-mode microplate reader is specially designed for medical, biological and pharmaceutical research and development institutions. It meets the diverse needs of various drug development and life science research programs with high-quality detection performance based on molecular biology, biochemistry and cytology.
Beyond the most basic absorbance, fluorescence and luminescence detection functions, high-performance fluorescence polarization (FP) and time-resolved fluorescence (TRF) modules are also available on select models, enabling researchers to conduct sophisticated assays with exceptional sensitivity and specificity.
Detection Modes Overview
Absorbance Detection
Powered by a xenon lamp light source and PD detector, the absorbance module covers a wavelength range of 200–1000 nm with 1 nm stepping. Wavelength accuracy reaches ±2 nm with repeatability (SD) of 0.2 nm. The measurement range spans 0–4 OD, and resolution is as fine as 0.0001 OD. At 450 nm, accuracy is ±(1.0%+0.003 Abs) in the 0.0–2.0 Abs range for the 96-well precision mode. A full 96-well plate can be read in under 15 seconds in fast mode.
Fluorescence Detection
The fluorescence module uses top-reading mode with a xenon lamp excitation source and PMT detector. Excitation wavelength covers 200–1000 nm while emission covers 270–850 nm. Three standard filter groups (EX485/EM530, EX523/EM564, EX624/EM692) are included, with additional wavelengths available as replaceable filters. Detection limit reaches 1 pM under optimized conditions, and the linear dynamic range extends over 6 logs.
Luminescence Detection
Equipped with a high-sensitivity PMT detector, the luminescence module achieves a detection limit of 15 amol/well. The linear dynamic range spans 6 logs with crosstalk ≤0.05%. Wavelength range covers 200–850 nm, and standard filters at 470 nm and 560 nm are provided.
Time-Resolved Fluorescence (TRF) — A400 / A500
Time-resolved fluorescence is based on lanthanide elements as dyes. After the excitation light is turned off, the emitted light can still be continuously expressed and released, thereby eliminating the interference of excitation light and scattered light. This technology offers high sensitivity, strong specificity, good stability, and short operation flow. It is ideally suited for ultra-micro analysis in biology and medicine, hormone detection, viral hepatitis marker detection, target cell marker detection, and drug screening. The TRF filter set (EX365/EM612) delivers a detection limit of 0.02 pM under optimized conditions.
Fluorescence Polarization (FP) — A500
The optimized optical path design of the Feyond-A500 combined with the performance of a fast-switching polarizer can effectively reduce detection deviation. FP is commonly used to detect interactions between small molecules and macromolecules, such as determination of drugs and hormones, tyrosine kinase detection, receptor/ligand research, protein/polypeptide interactions, DNA/protein interactions, and more. The FP filter set (EX485/EM530) achieves a detection limit of 5 mP.
Optical Specifications Summary
| Parameter | Feyond-A300 / A400 / A500 |
|---|---|
| Absorbance Light Source | Xenon lamp |
| Absorbance Detector | PD |
| Wavelength Range (Abs) | 200–1000 nm, 1 nm step |
| Wavelength Accuracy | ±2 nm |
| Measure Range | 0–4 OD |
| Resolution | 0.0001 OD |
| Reading Time (96-well) | Fast <15 s |
| Luminescence Detector | PMT |
| Luminescence Detection Limit | 15 amol/well |
| Fluorescence Excitation Range | 200–1000 nm |
| Fluorescence Emission Range | 270–850 nm |
| Fluorescence Detection Limit | 1 pM |
| TRF Detection Limit (A400/A500) | 0.02 pM |
| FP Detection Limit (A500) | 5 mP |
Instrument Configuration
The Feyond series supports 6- to 384-well microplates with three shaking modes (linear, annular, double annular). Incubation temperature ranges from RT+4°C to 45°C with temperature uniformity of ±0.5°C at 37°C. The instrument features a 10-inch capacitive touch screen with mouse support, Chinese/English software interface, and 10 GB data storage capacity. Network transmission via FTP allows test data reports to be uploaded to a PC server for remote access.
Connectivity includes 2 USB Type-A ports, 1 USB Type-B port, 1 Ethernet port, and an RS232 bus interface for injector connection. The optional MSS-2 automatic injector can be purchased together with the instrument or upgraded later, with dispensing volume of 5–1000 μL in 1 μL increments and injection speed of 125–500 μL/s.
Ordering Information
| Code | Product Description |
|---|---|
| 63-RS-19050-00 | Feyond-A300 microplate reader (multi-mode) |
| 63-RS-19060-00 | Feyond-A400 microplate reader (multi-mode) |
| 63-RS-19070-00 | Feyond-A500 microplate reader (multi-mode) |
| 63-RS-19011-01 | ReaderIt-II PC analysis software |
| 63-RS-19011-02 | u-Nano ultra-micro plate |
| 63-RS-19011-03 | ABS optical performance validation board |
| 63-RS-19011-04 | MSS-2 automatic injector |
| 63-RS-19051-01 | Lum-check standardization light microplate |
| 63-RS-19051-02 | 470-525 fluorescence filter (standard) |
| 63-RS-19051-03 | 523-564 fluorescence filter (standard) |
| 63-RS-19051-04 | 624-692 fluorescence filter (standard) |
| 63-RS-19051-41 | LUM-470 filter (standard) |
| 63-RS-19051-05 | 365-612 time-resolved fluorescence filter (standard) |